Pulse oximetry & capnography — use and pitfalls
SpO₂ / EtCO₂ · Critical Care / ICU
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Oximetry: two-wavelength light absorption across a pulsatile bed estimates functional SaO₂. Capnography: continuous expired CO₂ — the waveform confirms tracheal intubation, tracks ventilation, perfusion and circuit integrity.
Red flags
- Normal SpO₂ in a smoke-inhalation patient means nothing — measure COHb
- Loss of the capnograph trace in a ventilated patient is an emergency, not an artefact
Management
- Treat the patient, not the probe: check waveform quality before believing a bad number
- Oximetry errors: carboxyhaemoglobin reads falsely HIGH (CO poisoning — use a co-oximeter gas), methaemoglobin drives readings toward ~85%, poor perfusion/vasoconstriction, nail varnish, movement, dyes
- SpO₂ is blind to ventilation — a normal saturation on oxygen can hide severe hypercapnia; get a gas when in doubt
- Capnography: flat trace after intubation = oesophageal tube or no circulation until proven otherwise
- Falling EtCO₂: falling cardiac output/PE/hypovolaemia or circuit leak; rising: hypoventilation, malignant hyperthermia, bicarbonate
- Use continuous waveform capnography for EVERY intubation, transfer and cardiac arrest (trend quality of CPR; sudden rise = ROSC)
Oximetry measures oxygenation only; capnography guards the airway, the ventilation and the circulation.