Arterial line & blood pressure monitoring
Art line · Critical Care / ICU
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Continuous beat-to-beat pressure plus sampling access for the unstable patient. Radial first choice (collateral flow, accessible); femoral/brachial give more central pressures but brachial is an end-artery — treat thrombosis there as an emergency.
Red flags
- White/painful hand distal to a radial line — remove and assess perfusion immediately
- Accidental intra-arterial drug injection: leave the cannula, do not flush blindly — analgesia, anticoagulation discussion, vascular advice
Management
- Indications: shock/vasoactive infusions, frequent gases, large swings expected (induction, dissection control)
- Aseptic insertion; transducer zeroed at the phlebostatic axis (4th intercostal space, mid-axillary line); re-zero after moves
- Check damping: a flushed system should return crisply — over-damped (bubbles, clot, kinks) under-reads systolic; under-damped over-reads (“ringing”)
- Read the waveform: swing with ventilation suggests fluid-responsiveness; slow upstroke = poor contractility/stenosis; steep diastolic decay = vasodilatation
- Never trust one number: a “normal” MAP on rising vasopressors is a deteriorating patient; treat perfusion (lactate, urine, mentation), not the display
- Remove when no longer needed — daily line review; watch for distal ischaemia, and NEVER inject drugs into an arterial line
Zero it, damp-test it, and read the waveform — the shape carries as much information as the number.
Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.