Supraglottic airway (LMA) failure — rescue strategy
Airway rescue · Perioperative & Anaesthesia
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
A supraglottic airway (SGA) may fail to provide a patent airway, adequate ventilation or aspiration protection due to malposition, inadequate depth or patient factors; a structured rescue avoids hypoxia and escalation to a CICO situation.
Red flags
- Inadequate or absent capnography trace with poor chest rise
- Falling SpO2 despite optimisation
- Large audible leak with inadequate tidal volumes
- Regurgitation or gastric contents in the device
- Progression towards inability to ventilate by any means
Assessment
Differential diagnosis
- Malposition / inadequate seal (folded cuff, wrong size, down-folded epiglottis)
- Inadequate anaesthetic depth causing coughing, laryngospasm or straining
- High airway pressures / poor compliance (obesity, bronchospasm, pneumoperitoneum)
- Airway obstruction by the epiglottis or arytenoids
- Gastric insufflation and regurgitation with a first-generation device
- Surgical/positional displacement (shared airway, prone, head/neck surgery)
Investigations
- Capnography (confirming ventilation and detecting obstruction/leak)
- Tidal volumes, airway pressures and audible leak assessment
- SpO2 trend
- Direct visualisation or fibreoptic view through the SGA to assess position
- Chest rise and bilateral air entry
- Oropharyngeal leak pressure assessment
Management
- Give 100% oxygen and call for help if ventilation is inadequate
- Deepen anaesthesia if the problem is light anaesthesia (coughing, straining, laryngospasm)
- Optimise the device: chin lift/jaw thrust, reposition the head, adjust depth of insertion and reinflate the cuff to the correct pressure
- Check and correct the cuff volume/pressure (avoid over-inflation which worsens seal)
- Try an up–down manoeuvre or reinsert; consider a different size or a second-generation device
- If still failing, remove the SGA and provide 2-person face-mask ventilation with adjuncts
- Consider intubation — directly, via videolaryngoscope, or fibreoptically through the SGA as a conduit
- Insert a gastric tube (second-generation device) to decompress the stomach and reduce aspiration risk
- Limit to a maximum of ~2–3 SGA attempts before moving to a definitive plan
- If oxygenation cannot be maintained by any means, declare CICO and proceed to front-of-neck access
- Reassess suitability of SGA versus tracheal tube for the planned surgery
- Document the failure mode and successful rescue
Drugs & doses
Propofol 0.5–1 mg/kg IV bolus or increase volatile to deepen anaesthesia
Suxamethonium 1–1.5 mg/kg IV (or 0.1–0.5 mg/kg for laryngospasm) to facilitate ventilation/intubation
Rocuronium 0.6–1.2 mg/kg IV to facilitate intubation if proceeding to a tracheal tube
Oxygen 100% throughout
Salbutamol 2.5–5 mg nebulised if high pressures are due to bronchospasm
Disposition
Continue surgery once a reliable airway is achieved; if a definitive tracheal tube was required or hypoxia occurred, document and monitor in recovery, escalating to HDU/ICU if indicated.
Complications & pitfalls
- Hypoxaemia
- Gastric insufflation, regurgitation and aspiration
- Airway trauma and sore throat
- Laryngospasm during manipulation
- Escalation to CICO
Exam pearls
- Most SGA failures are positional or due to light anaesthesia — optimise both before abandoning the device
- A second-generation SGA improves seal and allows gastric drainage and intubation conduit use
- Correct cuff pressure (around 60 cmH2O) improves seal more reliably than maximal inflation
- The SGA can be a rescue conduit for fibreoptic intubation in a difficult airway
- Know when to stop — do not persist with repeated SGA attempts into a CICO scenario