Quantitative neuromuscular monitoring
TOF monitoring · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Objective (quantitative) monitoring of the train-of-four ratio to confirm full recovery before extubation, preventing residual paralysis and its airway and respiratory complications.
Assessment
Investigations
- Acceleromyograph/electromyograph at the adductor pollicis
- Train-of-four count and ratio
- Post-tetanic count for deep block
- SpO2 and clinical strength tests
- Baseline calibration before relaxant
Management
- Calibrate the monitor before giving the relaxant
- Use train-of-four count to time intubation and reversal
- Use post-tetanic count when no twitches are present
- Guide reversal-agent choice and dose by the depth measured
- Confirm a train-of-four ratio of at least 0.9 before extubation
- Do not rely on clinical signs alone (they are insensitive)
- Document the recovery ratio achieved
Drugs & doses
No drugs — monitoring technique
Targets: TOF ratio 0.9 or greater before extubation
Guides neostigmine vs sugammadex dosing
Disposition
Theatre / recovery; essential whenever relaxants are used.
A train-of-four ratio below 0.9 defines residual block and raises aspiration and hypoxaemia risk — confirm objectively (Miller).