Moderate

Perioperative management of cardiac implantable electronic devices

Perioperative assessment · Perioperative & Anaesthesia
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Patients with a pacemaker or implantable cardioverter-defibrillator (ICD) require a plan to manage electromagnetic interference (chiefly from monopole diathermy), inappropriate shocks and pacing inhibition. Management depends on device type, indication, pacing dependence and surgical site.

Red flags

  • Pacing inhibition causing bradycardia/asystole during diathermy in a dependent patient
  • Inappropriate ICD shock during surgery
  • Loss of capture or device malfunction
  • Haemodynamic instability with uncertain underlying rhythm

Assessment

Differential diagnosis

  • Pacemaker-dependent patient at risk of pacing inhibition by diathermy
  • ICD at risk of inappropriate shock from diathermy artefact
  • Device malfunction or lead failure
  • Rate-responsive pacing triggered by surgical/vibration stimuli
  • Intrinsic arrhythmia unrelated to the device

Investigations

  • Identify device type, manufacturer, indication, pacing dependence and last check (device identity card/interrogation)
  • Recent device interrogation report (ideally within 12 months, sooner for ICDs)
  • ECG to determine underlying rhythm and pacing dependence
  • Chest X-ray if device/lead details unavailable

Management

  • Determine device type, pacing dependence and whether the surgical field is above or below the umbilicus (interference risk higher above)
  • Obtain cardiology/physiology advice and a recent interrogation; plan reprogramming where indicated
  • For pacing-dependent patients with high interference risk, reprogramme to asynchronous mode or apply a magnet (pacemakers) per advice
  • For ICDs, suspend tachytherapy (deactivate shocks) by reprogramming or magnet before diathermy, with external defibrillation pads applied and continuous monitoring
  • Use bipolar diathermy where possible; if monopolar, use short bursts and position the return plate so current does not cross the device/leads
  • Maintain continuous ECG with the ability to see the underlying rhythm, and have external pacing/defibrillation immediately available
  • Reactivate ICD therapies and reprogramme/interrogate the device postoperatively before removing continuous monitoring
  • Document device management plan and postoperative check

Drugs & doses

Standby chronotrope for pacing failure — Isoprenaline infusion 0.5-10 microgram/min IV titrated, or Atropine 0.5-0.6 mg IV
Arrhythmia with deactivated ICD — treat per ALS; have external defibrillation ready
Avoid drugs provoking arrhythmia near a deactivated ICD; correct electrolytes (e.g. potassium to 4.0-4.5 mmol/L)

Disposition

Monitored recovery until the device is interrogated/reactivated; stable patients return to the ward and pacing-dependent/ICD patients may need continued cardiac monitoring per cardiology advice.

Complications & pitfalls

  • Asystole from pacing inhibition
  • Inappropriate ICD discharge
  • Device reset or reprogramming errors
  • Thermal/lead damage from diathermy current path
  • Failure to reactivate ICD postoperatively

Exam pearls

  • A magnet over most pacemakers triggers asynchronous pacing, but over an ICD it suspends shocks WITHOUT changing pacing — know the difference
  • Always apply external defibrillation pads and continuous monitoring when an ICD's tachytherapy is deactivated
  • Prefer bipolar diathermy; if monopolar is essential, use short bursts and keep the current path away from the device and leads
  • Confirm the ICD is reactivated and the device re-checked before the patient leaves monitored care
  • Below-umbilicus surgery with bipolar diathermy often needs no reprogramming, but confirm with physiology

Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.