Oxygen supply failure / anaesthetic machine failure
Equipment emergency · Perioperative & Anaesthesia
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Loss of the oxygen supply or anaesthetic machine function is a life-threatening emergency that can rapidly cause hypoxia; the priority is to guarantee oxygenation and ventilation using an independent backup while maintaining anaesthesia.
Red flags
- Low FiO2 alarm with falling SpO2
- Inability to ventilate or deliver oxygen via the machine
- Suspicion of crossed/contaminated pipeline gases
- Simultaneous failure affecting multiple theatres (hospital supply failure)
- Patient awareness risk if volatile delivery is lost without IV cover
Assessment
Differential diagnosis
- Pipeline failure / crossed pipelines / pressure loss
- Cylinder empty or not opened; regulator failure
- Ventilator or bellows failure / power failure
- Circuit disconnection, leak or soda-lime exhaustion/channelling
- Vaporiser or flowmeter fault
- Oxygen analyser/alarm malfunction masking a hypoxic mixture
Investigations
- Inspired oxygen concentration (FiO2) analyser reading
- Oxygen supply pressure gauges (pipeline and cylinder)
- Capnography and airway pressure/volume monitoring
- SpO2 trend
- Circuit integrity check (disconnection, leaks)
- Ventilator and power status indicators
Management
- Recognise the problem (low FiO2/supply pressure alarm, failure to ventilate) and call for help
- Turn ON the backup oxygen cylinder on the anaesthetic machine and check it is full
- Disconnect the pipeline supply if a crossed pipeline or pipeline fault is suspected
- Switch to a self-inflating bag (e.g. Ambu) with an independent oxygen cylinder to guarantee ventilation
- Maintain anaesthesia intravenously (TIVA) since volatile delivery may be lost
- Confirm oxygenation and ventilation with SpO2 and capnography
- Use minimal/low fresh-gas flow or manual ventilation to conserve cylinder oxygen while help arrives
- Call for additional cylinders and alert theatre coordinator, estates/engineering and other theatres
- Consider aborting/pausing surgery at a safe point if supply cannot be secured
- Identify and isolate the fault (machine, pipeline, hospital supply)
- Keep the patient monitored and anaesthetised throughout the transition
- Document and report the incident and quarantine faulty equipment
Drugs & doses
Propofol TCI or 100–200 micrograms/kg/min infusion to maintain anaesthesia (TIVA)
Remifentanil TCI / 0.05–0.25 micrograms/kg/min infusion for analgesia/anaesthesia
Midazolam 1–2 mg IV or ketamine 0.25–0.5 mg/kg IV as backup amnesia if infusions limited
Neuromuscular blocker as required to maintain relaxation
Oxygen from an independent full cylinder via self-inflating bag
Disposition
Complete or safely abandon surgery; continue in recovery/ICU as the patient's state dictates, report the incident, and remove faulty equipment from service until engineering clearance.
Complications & pitfalls
- Hypoxia and hypoxic organ injury
- Awareness under anaesthesia
- Hypercapnia from inadequate ventilation
- Delivery of a hypoxic gas mixture (crossed pipelines)
- Haemodynamic instability
Exam pearls
- A self-inflating bag with its own oxygen cylinder is the ultimate backup — have one in every theatre
- If in doubt, disconnect from the machine/pipeline and ventilate independently
- Maintain anaesthesia with TIVA when gas delivery is compromised to prevent awareness
- Oxygen failure warning devices (and the oxygen ratio/pressure interlock) are key safety features
- Perform and document the pre-use machine check to detect many faults before they cause harm
Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.