Inadvertent volatile anaesthetic overdose

Volatile overdose · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

An excessively high inspired volatile concentration causes profound hypotension, bradycardia and reduced cardiac output, and may be due to vaporiser error or filling fault. Reducing the agent and supporting the circulation reverse it.

Assessment

Investigations

  • Measured end-tidal volatile concentration (MAC)
  • Blood pressure, heart rate and cardiac output
  • Depth-of-anaesthesia monitor
  • Inspect vaporiser setting and filling
  • ECG

Management

  • Turn the vaporiser off or to minimum
  • Increase fresh gas flow to wash out the agent
  • Give 100 percent oxygen and support ventilation
  • Treat hypotension with fluids and vasopressors
  • Treat bradycardia
  • Check for vaporiser overfilling, tipping or wrong agent

Drugs & doses

Fluid bolus 250-500 mL crystalloid
Metaraminol 0.5 mg IV boluses or noradrenaline infusion
Atropine 0.5 mg IV for bradycardia

Disposition

Continue once stable; remove faulty vaporiser from service.

An unexpectedly deep, hypotensive patient with a high end-tidal agent reading points to vaporiser overdelivery — the agent analyser is the quickest check.

Disclaimer: Educational decision support for qualified clinicians and students — not a substitute for clinical judgement, local protocols or a full drug reference. Confirm every dose and contraindication against your formulary before prescribing.