Capnograph sampling faults

Monitoring safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Capnography hardware problems such as a blocked or water-logged sampling line, loose connectors or an exhausted absorber can mimic or mask clinical events. Systematic checks restore a reliable trace.

Assessment

Investigations

  • Inspect the sampling line for kinks and water
  • Check connector seating (sidestream vs mainstream)
  • Confirm the analyser is calibrated and zeroed
  • Soda lime colour (raised baseline suggests exhaustion)
  • Correlate with SpO2 and clinical signs

Management

  • Confirm whether the problem is the patient or the equipment
  • Check the sampling line for kinks, disconnection and water
  • Clear or replace a blocked sidestream line and water trap
  • Verify connectors and the airway adaptor are seated
  • Recalibrate or zero the analyser if the baseline is wrong
  • Replace exhausted soda lime if the baseline is raised
  • Revert to clinical assessment while troubleshooting

Drugs & doses

No drugs — monitoring procedure
Spare sampling lines and water traps to hand
Replace absorber if inspired CO2 rises

Disposition

Theatre; never ignore a lost trace without excluding disconnection.

A raised baseline indicates absorber exhaustion or rebreathing; always exclude true disconnection before blaming the monitor (Miller).