Breathing-circuit disconnection and leak during anaesthesia
Disconnection · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
Disconnection or a major leak in the breathing system causes loss of ventilation and is a leading cause of avoidable hypoxic harm. Capnography and a low-pressure/volume alarm give the earliest warning.
Assessment
Investigations
- Loss of capnograph trace and falling tidal volume
- Low airway pressure and disconnect alarms
- Systematic check of every circuit junction
- Reservoir bag movement
- SpO2 trend
Management
- Confirm the problem: flat capnograph with low-pressure alarm
- Give 100 percent oxygen and hand ventilate to localise the leak
- Check the commonest sites: catheter mount, filter, Y-piece, bag, APL valve and CO2 absorber seatings
- Reconnect or replace the faulty component
- If the leak cannot be found, use a self-inflating bag and alternative oxygen source
- Resume ventilation and confirm return of capnograph trace
Drugs & doses
100 percent oxygen
Self-inflating (Ambu-type) bag with reservoir if machine failure
Disposition
Continue case; report equipment fault.
A flat capnograph is a ventilation failure until proven otherwise — never silence the alarm without finding the cause (Assoc of Anaesthetists machine-check guidance).