Breathing-circuit disconnection and leak during anaesthesia

Disconnection · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Disconnection or a major leak in the breathing system causes loss of ventilation and is a leading cause of avoidable hypoxic harm. Capnography and a low-pressure/volume alarm give the earliest warning.

Assessment

Investigations

  • Loss of capnograph trace and falling tidal volume
  • Low airway pressure and disconnect alarms
  • Systematic check of every circuit junction
  • Reservoir bag movement
  • SpO2 trend

Management

  • Confirm the problem: flat capnograph with low-pressure alarm
  • Give 100 percent oxygen and hand ventilate to localise the leak
  • Check the commonest sites: catheter mount, filter, Y-piece, bag, APL valve and CO2 absorber seatings
  • Reconnect or replace the faulty component
  • If the leak cannot be found, use a self-inflating bag and alternative oxygen source
  • Resume ventilation and confirm return of capnograph trace

Drugs & doses

100 percent oxygen
Self-inflating (Ambu-type) bag with reservoir if machine failure

Disposition

Continue case; report equipment fault.

A flat capnograph is a ventilation failure until proven otherwise — never silence the alarm without finding the cause (Assoc of Anaesthetists machine-check guidance).