Breathing-circuit disconnection

Circuit emergency · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026

Bedside summary

Disconnection of the breathing system is a leading cause of avoidable critical hypoxia under anaesthesia. Capnography loss and a low-pressure alarm allow rapid recognition and correction.

Assessment

Investigations

  • Capnography (sudden loss of trace)
  • Airway pressure and disconnect alarm
  • SpO2 (falls later)
  • Ventilator minute volume
  • Visual trace of every connection

Management

  • Recognise the sudden loss of capnograph trace and alarm
  • Switch to hand ventilation to assess the circuit
  • If uncertain, ventilate with a self-inflating bag and oxygen
  • Trace the circuit from patient to machine for the break
  • Check the Y-piece, filter, sample line and APL valve
  • Reconnect securely and confirm the trace returns
  • Verify oxygenation and ventilation before continuing

Drugs & doses

No drugs — equipment emergency
Keep a self-inflating bag immediately available
Confirm return of a normal capnograph trace

Disposition

Theatre; investigate the cause before proceeding.

Loss of the capnograph trace with a low airway-pressure alarm is disconnection until proven otherwise — ventilate by hand first (AAGBI).