Bedside summary

Workstation ventilator alarms (high pressure, low pressure/disconnect, low minute volume, apnoea) warn of circuit and patient problems. A systematic response prevents hypoxia and barotrauma.

Assessment

Investigations

  • Airway pressure waveform and value
  • Expired minute and tidal volume
  • Capnography and SpO2
  • Inspect circuit, tube and patient
  • Ventilator settings and gas supply

Management

  • Note which alarm has triggered and its message
  • For high pressure, check for kinking, obstruction, bronchospasm or light anaesthesia
  • For low pressure or disconnect, trace the circuit for leaks
  • For low minute volume, check leaks, settings and compliance
  • For apnoea, confirm ventilation and the gas supply
  • Hand-ventilate if the cause is not quickly found
  • Correct the cause and confirm normal gas exchange

Drugs & doses

No drugs — equipment response
Self-inflating bag available as backup
Deepen anaesthesia or give a relaxant if the patient is fighting

Disposition

Theatre.

Treat a high-pressure alarm as obstruction, light anaesthesia or bronchospasm until excluded; hand-ventilate when in doubt (Morgan).