Anaesthetic machine check (AAGBI)
Machine safety · Anaesthesia & Perioperative
Physician-written · Verified by three sets of doctors · Reviewed November 2026
Bedside summary
A structured pre-use check of the anaesthetic workstation and breathing system performed before every operating list to detect faults that could cause hypoxia, awareness or barotrauma.
Assessment
Investigations
- Self-test per the machine manufacturer
- Oxygen analyser calibration (21% and 100%)
- Pipeline tug test and cylinder reserves
- Two-bag test of the breathing system and ventilator
- Capnography, vaporiser filling and scavenging checks
Management
- Confirm the machine is plugged in with a reserve oxygen cylinder
- Run the manufacturer self-test
- Check pipelines by tug test and verify gauge pressures
- Calibrate the oxygen analyser
- Confirm vaporisers are seated, filled and leak-free
- Perform the two-bag test on the circuit and ventilator
- Check scavenging, suction, airway equipment, monitors and alarms
Drugs & doses
No drugs — equipment procedure
Verify an alternative means of ventilation (self-inflating bag) is present
Record the check before the first case
Disposition
Theatre; do not start without a completed check.
Perform a machine and breathing-system check before every list and after any change; always have a self-inflating bag available (AAGBI machine check 2012).